Confirmed technical specifications| Dimensions | 800 × 450 × 1800 mm |
|---|
| Operating temperature | -40 °C to 70 °C |
|---|
| Display size | 9 in |
|---|
| Data storage | 16 GB |
|---|
| Sampling method | Pump |
|---|
- Documented measurement scope
- NO, NO₂, SO₂, NH₃, H₂S, O₃, benzene, toluene and xylene; the document does not assign every listed range to every gas
- Documented range options
- 0–2, 10, 20, 50, 100, 500, 1000 or 2000 ppm or mg/m³; gas-to-range selection requires project confirmation
- Measurement principle
- Differential ultraviolet absorption spectroscopy
- Sampling temperature
- -40 to 600 °C standard; 800–1300 °C configurations optional
- Sampling humidity
- 0–99% RH
- Sampling distance
- 30–40 m standard; over 70 m with optional vacuum pump
- Sampling flow
- 2 L/min standard; 4 L/min optional
- Signal output
- 4–20 mA, RS485 and multiple passive contacts; wireless or network transmission optional
- Enclosure
- Outdoor waterproof cabinet; explosion-protected configuration optional
- Power supply
- 220 V AC, 50 Hz